| 4–20 mA Current | Source and measure current; simulate a two-wire transmitter; provide loop power when required. | 4 mA = 0%; 12 mA = 50%; 20 mA = 100%. Industrial loops commonly use approximately 24 V DC supply power. | Choose a unit with at least 0–24 mA sourcing and measurement capability, fine adjustment, percentage display, and current-loop power output. | Pressure, temperature, flow, level, valve positioners, and process transmitters. | Check source-versus-measure functions, loop compliance voltage, burden voltage, fuse protection, and accuracy at low current values. |
| HART Communication | Support digital communication over the 4–20 mA loop while measuring or sourcing analog current. | HART uses frequency-shift keying with 1,200 Hz and 2,200 Hz tones superimposed on the analog signal; a 250 Ω loop resistance is commonly used. | Select a calibrator with a built-in HART modem, loop resistance control, device configuration access, and support for digital trim or range verification. | Smart pressure, temperature, flow, and level transmitters. | Verify compatibility with the required communication revision, device description files, loop resistance, power level, and multidrop limitations. |
| RTD Temperature | Measure and simulate resistance temperature detectors using two-, three-, and four-wire connections. | Pt100 is widely used; IEC 60751 defines the industrial platinum RTD reference relationship. Other common sensors include Pt1000, Ni100, and Ni120. | Choose support for the required sensor types, wiring methods, temperature units, custom resistance values, and low-resistance measurement stability. | Process temperature transmitters, heat-treatment systems, reactors, tanks, and utility equipment. | Confirm IEC 60751 or applicable curve selection, lead-wire compensation, excitation current, self-heating effects, and sensor conversion accuracy. |
| Thermocouple Temperature | Source and measure thermocouple millivolt signals with automatic or manual cold-junction compensation. | Common types include J, K, T, E, N, R, S, and B. Reference tables are maintained through international and national thermocouple standards. | Select the thermocouple types used on site, stable cold-junction compensation, insulated or grounded input support, and adequate millivolt resolution. | Furnaces, boilers, kilns, combustion systems, exhaust monitoring, and high-temperature process equipment. | Check reference-junction temperature, extension-wire compatibility, polarity, thermocouple plug type, noise rejection, and accuracy across the working range. |
| Pressure | Measure pressure and, where needed, generate pressure using an external hand pump or integrated pressure module. | Gauge, absolute, differential, and vacuum measurements are common. Typical field ranges may extend from low-pressure differential service to several hundred bar, depending on the process. | Match the sensor range to the application, preferably avoiding operation near the upper limit; select accuracy appropriate for the transmitter under test. | Pressure transmitters, switches, gauges, filtration systems, pumps, compressors, and custody-related checks. | Verify pressure type, overpressure protection, media compatibility, fittings, leak tightness, temperature effects, and whether an external reference is required. |
| Frequency and Pulse | Generate and measure frequency, pulse rate, duty cycle, or totalized pulse count where flow or speed instruments require it. | Pulse inputs are often specified by frequency, voltage level, threshold, and pulse width rather than by a universal process standard. | Choose adjustable frequency, amplitude, duty cycle, pulse count, and input-threshold measurement. | Flowmeters, tachometers, batching systems, and digital totalizers. | Confirm signal level, isolation, maximum frequency, minimum pulse width, and compatibility with the instrument input circuit. |
| Switch and Continuity Test | Test open/close state, trip point, reset point, and contact continuity. | Common for pressure, temperature, level, and flow switches; trip values are defined by the device and process specification. | Select adjustable ramping, switch-state detection, programmable trip limits, and clear pass/fail recording. | Alarm switches, shutdown devices, interlocks, and control panels. | Check contact wetting current, relay contact ratings, hysteresis measurement, ramp rate, and safe isolation from the process circuit. |